Zirconia Oxygen Analyser


Discover Fuji Electric ZKM Direct Insertion Zirconia Oxygen Analyser
analyser_enquiry

General Specifications


Application: Oxygen in noncombustible gas
Measuring method: Directly inserted type zirconia cell
Measuring range: 0 to 2 ··· setting range at option 2 in 50 vol% O2 (in 1 vol% O2 steps)
Repeatability: Within ±0.5%FS
Linearity: Within ±2%FS
Response time: Less than 4 sec, for a 90% step change
Warmup time: approx 10 min
Analog output: 4 to 20mA DC (allowable load resistance less than 500Ω) or 0 to 1V DC (output resistance more than 100Ω)
Power supply:
Rated voltage;
100 to 120V AC (operating voltage 90 to 132V AC)
200 to 240V AC (operating voltage 190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption: Maximum 240VA (Detector: approx. 200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA, Converter: approx. 20VA)



 

fuji electric e front runners logo

Fuji Electric ZKM In-situ Zirconia Oxygen Analyser


This oxygen analyser is deigned to provide relieable, fast responce continuously measure oxygen concentration in combustible exhaust gas of industrial boilers or furnaces, and is ideally suited for combustion control and management. The analyser system is comprised of the detector and converter coupled together as a complete system. Detector setting configuration includes the detector flow tube and zirconia detecting sensor. The flow tube is inserted directly into the gas and directs gas to the sensor for measurement. The converter (ZKM) is comprised of the signal processor, input/output and communications, display and system controls. The converter is equipped with advanced functionality such as performing the sensor diagnostics and sensor recovery function, so the detector can be used within long term stability.
Fuji Oxygen Gas Analyser

Features


1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the flue. Gas sampling functions such as a gas aspirator and a dehumidifier are not required.

2. Easy maintenance
The sensor is equipped with a detector and is easy to replace. By separating the detector and the flow guide tube, filter replacement is made easy.

3. More reliable than sensor diagnosis, sensor recoverable function
Depending on the concentration of the measured gas, the power of the sensor may deteriorate. The equipment includes sensor recovery function electronically, checking the deterioration status of the sensor, thereby giving high reliability and long-lasting stability.

4. Safe and secure
The system detects thermocouple break for heater control on the sensor side. The analyser features the safety functions of either isolating power supply to the detector, or isolating power via external contact input.

5. Easy operation
The operation and settings for the converter can be performed via the front panel.

Analysing air pollution is the first step toward preserving a clean atmospheric environment. FUJI developed the first infra-red gas analyser in Japan to use mass-flow sensors. Since then, we have supplied customers with various types of gas analysers to support environmental preservation and control efforts. These efforts include measurements of atmospheric pollution and detection of low density of SOx and NOx, generated by incinerating facilities and boilers. FUJI's gas analysers are commonly used to monitor the atmosphere to help maintain a cleaner natural environment





.
Fuji Probe for Gas Analyser


analyser_enquiry

Colour Brochure

One Download

ZKM - ZFK8 Colour Brochure (508 Kb)

Datasheet

2 Downloads

Gas Analyzer (ZRY) (137 Kb)
ZKM - ZFK8 Datasheet (299 Kb)

Instruction Manual

2 Downloads

ZRY Instruction Manual (180 Kb)
ZKM Instruction Manual (1883 Kb)
 
 
 
Home Page
Products
Transmitters
Flowmeters
Switches
Signal Conditioners
Indicators
Controllers
Recorders
Analysers
Oxygen
Gas
Gas (High Perf)
Control Valves
Repairs and Service
Information
Forum
Links
www.intergage.co.uk | Web Site Content Management